Testo 350 Analysis Box for Exhaust Gas Analysis Systems

SKU:

0632-3510 Testo 350 Analysis Box for Exhaust Gas Analysis Systems
Testo logo

Ideal for professional flue gas analysis and industrial emissions measurement on industrial engines, burners, gas turbines and in thermal processes: the testo 350 flue gas analyzer performs a varie...

Contact us for pricing

This product requires a custom quote

In Stock

Quantity

Description

Ideal for professional flue gas analysis and industrial emissions measurement on industrial engines, burners, gas turbines and in thermal processes: the testo 350 flue gas analyzer performs a variety of measuring and analysis tasks, impresses over the long term thanks to its heavy-duty industrial design and is also suitable for complex data acquisition.

 

The testo 350 flue gas analyzer comprises two units

  • The testo 350 Control Unit (separate product) with a clear graphic display, allows you to control emissions measurement. Operation is really easy: burner, gas turbine, engine and user-defined applications are available for selection in the menu. The Control Unit enables remote control of the analyzer unit even when there is a spatial separation of the flue gas pipe and the setting location – particularly recommended for emissions measurements on large plants.
  • The testo 350 analyzer unit is just what you need for carrying out emissions measurement, because it contains all the sensor technology and electronics. The testo 350 analyzer unit includes an O2 gas sensor as standard, but at least one additional sensor must be connected for commissioning (can be used with a maximum of 6 sensors). When connecting the optional sensors, you can choose between gas sensors for CO, CO2, NO, NO2, SO2, H2S or CxHy.

    The measuring range extension enables unrestricted measurements to be carried out, even when there are high gas concentrations. In order to protect the sensor technology, the measuring range extension (dilution) is automatically activated when there are unexpectedly high gas concentrations. The measuring range of a selected sensor can be extended by a specific factor. The service opening on the underside of the instrument allows fast access to all relevant serviceable and wearing parts, such as pumps and filters, enabling users to clean or replace these. In addition, the testo 350 has numerous instrument diagnosis functions. Instrument notifications are output in cleartext, making them easy to understand. The current status of the flue gas analyzer is constantly displayed.

     

    Using the testo 350 analyzer unit

    The analyzer unit can be operated with the Control Unit and it can also be used in direct combination with a PC/notebook and the easyEmission software (via USB, Bluetooth® 2.0 or CANCase). After programming, the analyzer unit can carry out measurements and save the results autonomously. It is also possible to display or transmit the measurement data from the analyzer unit to the Control Unit or software.

    To send the measurement report straight away, you can also use our free App which turns your Android smartphone or tablet into a display device for the analyzer unit.

    Additional Details

    testo 350 analyzer unit, equipped with O 2 , including differential pressure sensor, temperature probe input type K NiCr-Ni and type S, Pt10Rh-Pt, Testo data bus connection, rechargeable battery, integrated combustion air probe (NTC), trigger input, measurement data memory, USB interface, carrying strap kit and test protocol. Please note: The testo 350 analyzer unit can only be used in combination with the testo 350 Control Unit or the "easyEmission" software. These items are available separately.


Additional Information
No Additional Information

Product dimensions are not available.


Specifications
Absolute Pressure

Measuring range

600 to +1150 hPa

Accuracy

±10 hPa

Resolution

1 hPa

Butane (C₄H₁₀)

Measuring range

100 to 18000 ppm

Accuracy

< 400 ppm (100 to 4000 ppm) / 4000 ppm)

Resolution

10 ppm

Min. O₂ requirement in flue gas

2 % + (6.5 x mV butane)

Reaction time t₉₀

< 40 s

Response factor

2

footnote_

Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.

Differential Pressure - Piezoresistive

Measuring range

-200 to +200 hPa / -40 to +40 hPa

Accuracy

±1.5 % of mv (+3 to +40 hPa) / ±0.5 hPa (-49.9 to +49.9 hPa) / ±1.5 % of mv (-200 to -50 hPa) / ±0.03 hPa (-2.99 to +2.99 hPa) / ±1.5 % of mv (-40 to -3 hPa) / ±1.5 % of mv (+50 to +200 hPa)

Resolution

0.01 hPa (-40 to +40 hPa) / 0.1 hPa (-200 to +200 hPa)

Flue gas CO (with H₂-compensation)

Measuring range

0 to +10000 ppm

Accuracy

±10 ppm (0 to +199 ppm) / ±5 % of mv (+200 to +2000 ppm) / ±10 % of mv (+2001 to +10000 ppm)

Resolution

1 ppm (0 to +10000 ppm)

Reaction time t₉₀

40 s

footnote_

H₂ only as an indicator

Flue gas CO₂ - NDIR

Measuring range

0 to +50 Vol.%

Accuracy

±0.5 Vol.% + 1.5 % of mv (>25 to 50 Vol.%) / ±0.3 Vol.% + 1 % of mv (0 to 25 Vol.%)

Resolution

0.01 Vol.% (0 to 25 Vol.%) / 0.1 Vol.% (> 25 Vol.%)

Reaction time t₉₀

10 s

Flue gas CO₂ calculation (calculated from O₂)

Measuring range

0 to CO₂ max

Accuracy

calculated from O₂ ±0.2 Vol.%

Resolution

0.01 Vol.%

Flue gas COlow (with H₂-compensation)

Measuring range

0 to 500 ppm

Accuracy

±5 % of mv (+40 to +500 ppm) / ±2 ppm (0 to +39.9 ppm)

Resolution

0.1 ppm (0 to +500 ppm)

Reaction time t₉₀

40 s

footnote_

H₂ only as an indicator

Flue gas degree of effectivity, Eta (calculated)

Measuring range

0 to +120 %

Resolution

0.1 % (0 to +120 %)

Flue gas dewpoint (calculated)

Measuring range

0 to +99.9 °Ctd

Resolution

0.1 °Ctd (0 to +99.9 °Ctd)

Flue gas H₂S

Measuring range

0 to +300 ppm

Accuracy

±5 % of mv (+40 to +300 ppm) / ±2 ppm (0 to +39.9 ppm)

Resolution

0.1 ppm (0 to +300 ppm)

Reaction time t₉₀

35 s

Flue gas loss (calculated)

Measuring range

0 to +99.9 %

Resolution

0.1 % (0 to +99.9 %)

Flue gas NO

Measuring range

0 to +4000 ppm

Accuracy

±5 % of mv (+100 to +1999 ppm) / ±10 % of mv (+2000 to +4000 ppm) / ±5 ppm (0 to +99 ppm)

Resolution

1 ppm (0 to +4000 ppm)

Reaction time t₉₀

30 s

Flue gas NO₂

Measuring range

0 to +500 ppm

Accuracy

±5 % of mv (+100 to +500 ppm) / ±5 ppm (0 to +99.9 ppm)

Resolution

0.1 ppm (0 to +500 ppm)

Reaction time t₉₀

40 s

Flue gas Nolow

Measuring range

0 to +300 ppm

Accuracy

±5 % of mv (+40 to +300 ppm) / ±2 ppm (0 to +39.9 ppm)

Resolution

±0.1 ppm (0 to +300 ppm)

Reaction time t₉₀

30 s

Flue gas O₂

Measuring range

0 to +25 Vol.%

Accuracy

±0.8% of fsv (0 to +25 Vol.%)

Resolution

0.01 Vol.% (0 to +25 Vol.%)

Reaction time t₉₀

20 s (t₉₅)

Flue gas SO₂

Measuring range

0 to +5000 ppm

Accuracy

±10 % of mv (+2001 to +5000 ppm) / ±5 % of mv (+100 to +2000 ppm) / ±5 ppm (0 to +99 ppm)

Resolution

1 ppm (0 to +5000 ppm)

Reaction time t₉₀

30 s

Flue gas SO₂ low

Measuring range

0 to 200 ppm

Accuracy

±5 (40 to 200 ppm) / ±2 ppm (0 to 39.9 ppm)

Resolution

0.1 ppm

Reaction time t₉₀

30 s

footnote_

In combination with the special SO2low gas sampling probe 0563 1251 or the industrial heated probe kit 0600 7630

General technical data

Weight

4800 g

Dimensions

330 x 128 x 438 mm

Operating temperature

-5 to +45 °C

Product-/housing material

ABS

Protection class

IP40

Connectable probes

1 x combustion air temperature probe; 1 x flue gas probe; 1 x temperature probe; 1 x differential pressure

Product colour

Black

Battery life

5 h

Battery type

lithium batteries

DC electric tension input

11V to 40V

Power supply

AC mains unit 100V to 240V (50 to 60 Hz)

Dewpoint calculation

0 °Ctd; 99 °Ctd

Hose length

16.2 m (corresponds to 5 probe hose extensions)

Max. dust load

20 g/m³ dust in flue gas

Max. humidity load

70 °C dewpoint temperature

Max. neg. press./flue gas

-300 mbar

Max. pos. press./flue gas

50 mbar

Pump flow

1 l/min. with through-flow monitoring

Trigger input

Voltage 5 to 12 Volt (rising or trailing edge) pulse width > 1 sec load: 5 V/max, 5 mA, 12 V/max. 40 mA

Maximum memory

250,000 readings

Interface

data bus; Bluetooth®; USB; IR/IRDA interface; Mains connection; trigger input; Differential Pressure; probe input/ probe inputs; DC-voltage input; dilution gas inlet for measurement range extension

Storage temperature

-20 to +50 °C

footnote_

Methane (CH₄)

Measuring range

100 to 40000 ppm

Accuracy

< 400 ppm (100 to 4000 ppm) / 4000 ppm)

Resolution

10 ppm

Min. O₂ requirement in flue gas

2 % + (2 x mV methane)

Reaction time t₉₀

< 40 s

Response factor

1

footnote_

Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.

Propane (C₃H₈)

Measuring range

100 to 21000 ppm

Accuracy

< 400 ppm (100 to 4000 ppm) / 4000 ppm)

Resolution

10 ppm

Min. O₂ requirement in flue gas

2 % + (5 x mV propane)

Reaction time t₉₀

< 40 s

Response factor

1.5

footnote_

Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.

Temperature - NTC

Measuring range

-20 to +50 °C

Accuracy

±0.2 °C (-10 to +50 °C)

Resolution

0.1 °C (-20 to +50 °C)

footnote_

Ambient temperature probe

Temperature - TC Type K (NiCr-Ni)

Measuring range

-200 to +1370 °C

Accuracy

±1 °C (-200 to -100.1 °C) / ±1 °C (200.1 to +1370 °C) / ±0.4 °C (-100 to +200 °C)

Resolution

0.1 °C (-200 to +1370 °C)

Temperature - TC Type S (Pt10Rh-Pt)

Measuring range

0 to +1760 °C

Accuracy

±1 °C (0 to +1760 °C)

Resolution

0.1 °C (0 to +1760 °C)

Velocity / Volume flow

Measuring range

0 to +40 m/s

Resolution

0.1 m/s (0 to +40 m/s)

Ideal for professional flue gas analysis and industrial emissions measurement on industrial engines, burners, gas turbines and in thermal processes: the testo 350 flue gas analyzer performs a variety of measuring and analysis tasks, impresses over the long term thanks to its heavy-duty industrial design and is also suitable for complex data acquisition.

 

The testo 350 flue gas analyzer comprises two units

  • The testo 350 Control Unit (separate product) with a clear graphic display, allows you to control emissions measurement. Operation is really easy: burner, gas turbine, engine and user-defined applications are available for selection in the menu. The Control Unit enables remote control of the analyzer unit even when there is a spatial separation of the flue gas pipe and the setting location – particularly recommended for emissions measurements on large plants.
  • The testo 350 analyzer unit is just what you need for carrying out emissions measurement, because it contains all the sensor technology and electronics. The testo 350 analyzer unit includes an O2 gas sensor as standard, but at least one additional sensor must be connected for commissioning (can be used with a maximum of 6 sensors). When connecting the optional sensors, you can choose between gas sensors for CO, CO2, NO, NO2, SO2, H2S or CxHy.

    The measuring range extension enables unrestricted measurements to be carried out, even when there are high gas concentrations. In order to protect the sensor technology, the measuring range extension (dilution) is automatically activated when there are unexpectedly high gas concentrations. The measuring range of a selected sensor can be extended by a specific factor. The service opening on the underside of the instrument allows fast access to all relevant serviceable and wearing parts, such as pumps and filters, enabling users to clean or replace these. In addition, the testo 350 has numerous instrument diagnosis functions. Instrument notifications are output in cleartext, making them easy to understand. The current status of the flue gas analyzer is constantly displayed.

     

    Using the testo 350 analyzer unit

    The analyzer unit can be operated with the Control Unit and it can also be used in direct combination with a PC/notebook and the easyEmission software (via USB, Bluetooth® 2.0 or CANCase). After programming, the analyzer unit can carry out measurements and save the results autonomously. It is also possible to display or transmit the measurement data from the analyzer unit to the Control Unit or software.

    To send the measurement report straight away, you can also use our free App which turns your Android smartphone or tablet into a display device for the analyzer unit.

    Additional Details

    testo 350 analyzer unit, equipped with O 2 , including differential pressure sensor, temperature probe input type K NiCr-Ni and type S, Pt10Rh-Pt, Testo data bus connection, rechargeable battery, integrated combustion air probe (NTC), trigger input, measurement data memory, USB interface, carrying strap kit and test protocol. Please note: The testo 350 analyzer unit can only be used in combination with the testo 350 Control Unit or the "easyEmission" software. These items are available separately.