Ideal for professional flue gas analysis and industrial emissions measurement on industrial engines, burners, gas turbines and in thermal processes: the testo 350 flue gas analyzer performs a variety of measuring and analysis tasks, impresses over the long term thanks to its heavy-duty industrial design and is also suitable for complex data acquisition.
testo 350 analyzer unit, equipped with O 2 , including differential pressure sensor, temperature probe input type K NiCr-Ni and type S, Pt10Rh-Pt, Testo data bus connection, rechargeable battery, integrated combustion air probe (NTC), trigger input, measurement data memory, USB interface, carrying strap kit and test protocol. Please note: The testo 350 analyzer unit can only be used in combination with the testo 350 Control Unit or the "easyEmission" software. These items are available separately.
Length
33 cm
Width
12.8 cm
Height
43.8 cm
Weight
4.8 kg
Measuring range
600 to +1150 hPa
Accuracy
±10 hPa
Resolution
1 hPa
Measuring range
100 to 18000 ppm
Accuracy
< 400 ppm (100 to 4000 ppm) / 4000 ppm)
Resolution
10 ppm
Min. O₂ requirement in flue gas
2 % + (6.5 x mV butane)
Reaction time t₉₀
< 40 s
Response factor
2
footnote_
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
Measuring range
-200 to +200 hPa / -40 to +40 hPa
Accuracy
±1.5 % of mv (+3 to +40 hPa) / ±0.5 hPa (-49.9 to +49.9 hPa) / ±1.5 % of mv (-200 to -50 hPa) / ±0.03 hPa (-2.99 to +2.99 hPa) / ±1.5 % of mv (-40 to -3 hPa) / ±1.5 % of mv (+50 to +200 hPa)
Resolution
0.01 hPa (-40 to +40 hPa) / 0.1 hPa (-200 to +200 hPa)
Measuring range
0 to +10000 ppm
Accuracy
±10 ppm (0 to +199 ppm) / ±5 % of mv (+200 to +2000 ppm) / ±10 % of mv (+2001 to +10000 ppm)
Resolution
1 ppm (0 to +10000 ppm)
Reaction time t₉₀
40 s
footnote_
H₂ only as an indicator
Measuring range
0 to +50 Vol.%
Accuracy
±0.5 Vol.% + 1.5 % of mv (>25 to 50 Vol.%) / ±0.3 Vol.% + 1 % of mv (0 to 25 Vol.%)
Resolution
0.01 Vol.% (0 to 25 Vol.%) / 0.1 Vol.% (> 25 Vol.%)
Reaction time t₉₀
10 s
Measuring range
0 to CO₂ max
Accuracy
calculated from O₂ ±0.2 Vol.%
Resolution
0.01 Vol.%
Measuring range
0 to 500 ppm
Accuracy
±5 % of mv (+40 to +500 ppm) / ±2 ppm (0 to +39.9 ppm)
Resolution
0.1 ppm (0 to +500 ppm)
Reaction time t₉₀
40 s
footnote_
H₂ only as an indicator
Measuring range
0 to +120 %
Resolution
0.1 % (0 to +120 %)
Measuring range
0 to +99.9 °Ctd
Resolution
0.1 °Ctd (0 to +99.9 °Ctd)
Measuring range
0 to +300 ppm
Accuracy
±5 % of mv (+40 to +300 ppm) / ±2 ppm (0 to +39.9 ppm)
Resolution
0.1 ppm (0 to +300 ppm)
Reaction time t₉₀
35 s
Measuring range
0 to +99.9 %
Resolution
0.1 % (0 to +99.9 %)
Measuring range
0 to +4000 ppm
Accuracy
±5 % of mv (+100 to +1999 ppm) / ±10 % of mv (+2000 to +4000 ppm) / ±5 ppm (0 to +99 ppm)
Resolution
1 ppm (0 to +4000 ppm)
Reaction time t₉₀
30 s
Measuring range
0 to +500 ppm
Accuracy
±5 % of mv (+100 to +500 ppm) / ±5 ppm (0 to +99.9 ppm)
Resolution
0.1 ppm (0 to +500 ppm)
Reaction time t₉₀
40 s
Measuring range
0 to +300 ppm
Accuracy
±5 % of mv (+40 to +300 ppm) / ±2 ppm (0 to +39.9 ppm)
Resolution
±0.1 ppm (0 to +300 ppm)
Reaction time t₉₀
30 s
Measuring range
0 to +25 Vol.%
Accuracy
±0.8% of fsv (0 to +25 Vol.%)
Resolution
0.01 Vol.% (0 to +25 Vol.%)
Reaction time t₉₀
20 s (t₉₅)
Measuring range
0 to +5000 ppm
Accuracy
±10 % of mv (+2001 to +5000 ppm) / ±5 % of mv (+100 to +2000 ppm) / ±5 ppm (0 to +99 ppm)
Resolution
1 ppm (0 to +5000 ppm)
Reaction time t₉₀
30 s
Measuring range
0 to 200 ppm
Accuracy
±5 (40 to 200 ppm) / ±2 ppm (0 to 39.9 ppm)
Resolution
0.1 ppm
Reaction time t₉₀
30 s
footnote_
In combination with the special SO2low gas sampling probe 0563 1251 or the industrial heated probe kit 0600 7630
Weight
4800 g
Dimensions
330 x 128 x 438 mm
Operating temperature
-5 to +45 °C
Product-/housing material
ABS
Protection class
IP40
Connectable probes
1 x combustion air temperature probe; 1 x flue gas probe; 1 x temperature probe; 1 x differential pressure
Product colour
Black
Battery life
5 h
Battery type
lithium batteries
DC electric tension input
11V to 40V
Power supply
AC mains unit 100V to 240V (50 to 60 Hz)
Dewpoint calculation
0 °Ctd; 99 °Ctd
Hose length
16.2 m (corresponds to 5 probe hose extensions)
Max. dust load
20 g/m³ dust in flue gas
Max. humidity load
70 °C dewpoint temperature
Max. neg. press./flue gas
-300 mbar
Max. pos. press./flue gas
50 mbar
Pump flow
1 l/min. with through-flow monitoring
Trigger input
Voltage 5 to 12 Volt (rising or trailing edge) pulse width > 1 sec load: 5 V/max, 5 mA, 12 V/max. 40 mA
Maximum memory
250,000 readings
Interface
data bus; Bluetooth®; USB; IR/IRDA interface; Mains connection; trigger input; Differential Pressure; probe input/ probe inputs; DC-voltage input; dilution gas inlet for measurement range extension
Storage temperature
-20 to +50 °C
Measuring range
100 to 40000 ppm
Accuracy
< 400 ppm (100 to 4000 ppm) / 4000 ppm)
Resolution
10 ppm
Min. O₂ requirement in flue gas
2 % + (2 x mV methane)
Reaction time t₉₀
< 40 s
Response factor
1
footnote_
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
Measuring range
100 to 21000 ppm
Accuracy
< 400 ppm (100 to 4000 ppm) / 4000 ppm)
Resolution
10 ppm
Min. O₂ requirement in flue gas
2 % + (5 x mV propane)
Reaction time t₉₀
< 40 s
Response factor
1.5
footnote_
Lower explosion limit (LEL) must be adhered to. The HC sensor is adjusted to methane ex-works. It can be adjusted to a different gas (propane or butane) by the user.
Measuring range
-20 to +50 °C
Accuracy
±0.2 °C (-10 to +50 °C)
Resolution
0.1 °C (-20 to +50 °C)
footnote_
Ambient temperature probe
Measuring range
-200 to +1370 °C
Accuracy
±1 °C (-200 to -100.1 °C) / ±1 °C (200.1 to +1370 °C) / ±0.4 °C (-100 to +200 °C)
Resolution
0.1 °C (-200 to +1370 °C)
Measuring range
0 to +1760 °C
Accuracy
±1 °C (0 to +1760 °C)
Resolution
0.1 °C (0 to +1760 °C)
Measuring range
0 to +40 m/s
Resolution
0.1 m/s (0 to +40 m/s)
Ideal for professional flue gas analysis and industrial emissions measurement on industrial engines, burners, gas turbines and in thermal processes: the testo 350 flue gas analyzer performs a variety of measuring and analysis tasks, impresses over the long term thanks to its heavy-duty industrial design and is also suitable for complex data acquisition.
testo 350 analyzer unit, equipped with O 2 , including differential pressure sensor, temperature probe input type K NiCr-Ni and type S, Pt10Rh-Pt, Testo data bus connection, rechargeable battery, integrated combustion air probe (NTC), trigger input, measurement data memory, USB interface, carrying strap kit and test protocol. Please note: The testo 350 analyzer unit can only be used in combination with the testo 350 Control Unit or the "easyEmission" software. These items are available separately.